  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 96
85
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
Unexpected Messages and Observations
MESSAGE
POSSIBLE
CAUSE
SOLUTION
Apply Blood Bedside unit is
ready for blood
application.
Obtain a drop of capillary
blood and apply it to center
of pink test square.
If white pad becomes satu-
rated, you have applied
too much blood. Repeat
application with new test
strip.
If white patches or streaks
appear on confirmation
dot, you have not applied
enough blood. Repeat
application with new test
strip.
Bedside Unit
Disabled
Bedside unit
deleted from
system or
unassigned.
Contact system
administrator for current
status of bedside unit.
Bedside Unit
Needs
Configuration
Initial data transfer
is required. Or,
configuration is
corrupted.
Transfer data.