Apply control solution
Shake the control solution vial well.
Remove the cap.
Squeeze the vial and discard the first drop and
wipe off the dispenser tip with a clean tissue to
obtain an accurate result.
Squeeze the vial again to get a small hanging drop.
Apply the hanging drop b to the top edge area of
the OneTouch
®
Ultra Test Strip (where it meets the
narrow channel) until the drop is drawn into the
OneTouch
®
Ultra Test Strip, the confirmation
window is filled, and the meter display begins to
count down.
Read test results
The meter will count down from N to A and the
control solution test result will appear c.
Compare the result with the range printed on the
OneTouch
®
Ultra Test Strip vial label. The result
should fall within this range.
18 GETTING YOUR InDuo™ SYSTEM READY FOR USE